Choosing a Binding Assay

Choosing a Peptide Binding Assay

There is no single best method for measuring peptide binding. The right choice depends on what you need to know, how much material you have, whether you can label or immobilize a partner, and how tight the interaction is expected to be. This page compares the methods researchers most often use for peptides, with the practical trade-offs that are easy to discover too late.

Quick comparison

MethodWhat you getLabel or surface neededTypical affinity windowMaterial needed
SPR
Surface plasmon resonance
ka, kd and KD in real time One partner immobilized on a chip Very wide, pM to mM with care Low; microgram quantities
BLI
Biolayer interferometry
ka, kd and KD in real time One partner on a biosensor tip Roughly nM to high µM Low, but higher sample volume per well
ITC
Isothermal titration calorimetry
KD, stoichiometry and binding enthalpy None; free solution Roughly nM to tens of µM High; the main limitation
FP / FA
Fluorescence polarization
KD, and IC50 in competition mode Small fluorescent tracer peptide Bounded below by the tracer concentration Low; plate format
MST
Microscale thermophoresis
KD Fluorescent label, or intrinsic fluorescence Wide, pM to mM Very low volumes
ELISA
Peptide-coated plate
Relative binding, EC50 Peptide coated or captured; detection antibody Semi-quantitative Low; very high throughput
Pull-down
Biotinylated peptide and beads
Which proteins bind; usually qualitative Biotin tag and streptavidin support Not an affinity measurement Moderate
Ligand-observed NMR
STD, CPMG, WaterLOGSY
Binding yes/no, and binding epitope on the ligand None, but needs a spectrometer and expertise Best for weak binding, µM to mM High peptide, moderate protein
Affinity windows are indicative. The usable range for any method depends on the specific system, instrument and buffer.

How to narrow the choice

Do you need kinetics, or just affinity?

If the question is whether a peptide binds fast and lets go slowly — which matters for inhibitor work — you need a real-time method: SPR or BLI. If a single equilibrium constant answers the question, solution methods such as ITC, FP or MST avoid the complications of a surface.

Can you immobilize or label?

SPR, BLI, ELISA and pull-downs all require one partner attached to something. FP and MST require a fluorophore. ITC and NMR require neither, which is why they are so useful as orthogonal confirmation.

How much protein do you have?

ITC is the most informative single experiment and the most demanding: it typically needs hundreds of microliters at micromolar concentrations, and a matched buffer for both partners. If protein is scarce, it is usually the method you drop.

How weak might the interaction be?

Many isolated peptides bind in the high micromolar to millimolar range. FP struggles there, ITC runs out of window, and surface methods suffer from bulk refractive index effects. Ligand-observed NMR is built for exactly this range.

Method notes and characteristic artifacts

SPR

For peptides, which partner you immobilize matters more than anything else. SPR responds to mass arriving at the surface, so flowing a 2 kDa peptide over an immobilized protein gives a small signal. Immobilizing the peptide and flowing the protein usually gives a far better response.

The common artifacts are avidity and non-specific binding. If a dimeric or multivalent protein binds a densely coated peptide surface, the apparent affinity will be tighter than the true 1:1 value and the dissociation will look artificially slow. Keeping surface density low and checking that the measured KD does not change with density is the standard control. Peptides rich in arginine and lysine also bind carboxymethylated dextran surfaces non-specifically; a reference surface and a buffer with modest salt help.

BLI

BLI is more tolerant of crude or coloured samples than SPR and is easier to run in plate format, which makes it attractive for screening panels of peptides. It is generally less sensitive for small analytes and less able to resolve very fast kinetics. The same avidity considerations as SPR apply. Because the sensor moves between wells, evaporation and well-to-well variation deserve attention in long runs.

ITC

ITC is the closest thing to a direct measurement: no label, no surface, and it reports stoichiometry and enthalpy alongside affinity. Stoichiometry is a useful sanity check that is unavailable in most other formats. Its weaknesses are appetite for material and sensitivity to buffer mismatch; dialysing both partners into the same buffer is not optional, and heats of dilution must be measured and subtracted. Peptides stored as TFA salts can produce a buffer artifact, so counter-ion exchange is worth considering.

Fluorescence polarization

FP works when a small labeled peptide tumbles quickly in free solution and much more slowly once bound to a larger partner, so the size difference between tracer and target needs to be substantial. It shines in competition mode: establish a labeled tracer and target pair once, then rank unlabeled competitors cheaply in a plate.

The measurable affinity is limited by the tracer concentration, so a direct KD far below the tracer concentration cannot be determined accurately. Fluorophore choice is not neutral either: a hydrophobic dye can contribute to binding or cause the tracer to stick to surfaces.

MST

MST needs very little material and tolerates complex buffers, including serum and lysate in many cases. It is a good choice when the sample is precious or messy. Results are sensitive to aggregation and to adsorption on the capillary, so the controls that matter are a check that the fluorescence of the labeled partner does not change on binding, and visual inspection of the capillary scan.

ELISA and pull-down

These are discovery and screening tools rather than affinity measurements. A peptide ELISA is excellent for ranking a library or checking that an antibody recognizes its antigen, but a plate-bound peptide is multivalent and the readout usually involves an avidity-prone secondary antibody, so numbers should be treated as relative. A biotinylated-peptide pull-down from lysate answers "what binds this motif" and is a good partner to mass spectrometry; it says little about affinity.

Confirm in a second format

Methods fail in different ways. A surface method can report avidity as affinity; a fluorescence method can report dye behaviour as binding; a calorimetry experiment can report a buffer mismatch as heat. Two methods that disagree are telling you something useful. Two that agree are worth far more than one that is very precise.

What to decide before you order peptides

  • Which partner is immobilized or labeled, because that determines whether the peptide needs biotin, a fluorophore or a free terminus.
  • Where the tag goes, N-terminal or C-terminal, and whether a spacer is needed between the tag and the binding motif.
  • Which controls you will synthesize at the same time: a scrambled sequence, a point mutant that should abolish binding, and for modification studies a matched unmodified peptide.
  • Purity and quantity, including whether you need the counter-ion exchanged and the net peptide content measured for accurate concentrations.

Ordering the controls in the same batch as the test peptides is worth doing. They are made with the same chemistry and analyzed the same way, which removes one source of doubt when the result is surprising.